Menu
Skip to main content
How Can We Help?

Search for answers or browse our help center.

Print

How to add patient stop switch

REPETITIVE MOTION TABLE:

This is the current wiring. Insert the switch leads into this connector.

Screen_Shot_2018-05-29_at_8.33.11_AM.png

Red goes in the 6th position (top row). Black goes in the 2nd position (bottom row).

Screen_Shot_2018-05-29_at_8.34.26_AM.png

Mount the jack to the skirt by drilling a 3/8″ hole on the surface of the base enclosure that includes the power switch. Insert the jack into the hole from the inside and secure it by threading the nut on the outside.

Screen_Shot_2018-05-29_at_8.36.32_AM.png

This holder can be mounted to the underside of the cushion to store the switch when it is not being used.

Screen_Shot_2018-05-29_at_8.35.49_AM.png