Menu
Skip to main content
How Can We Help?

Search for answers or browse our help center.

Print

Replace Corded Pedal

BASE – Replace Corded Pedal

PARTS

corded pedal, crimp connectors on leads

Heyco cord restraint wire ties (3)

  1. Use the Height pedal (if it functions) to elevate the table.

If the pedal does not function the height of the table might still be enough that, if the steps 3 and 4 are done, you will be able to access the area where the wire is connected inside the base. If not, the table will need to be accessed from the bottom.

  1. Unplug the power cord from the wall receptacle.
  1. Remove 6 screws that secure the inside section of plastic skirting to the base.
  1. Use tape or string to hold the skirt up so that you can access the wiring.
  1. Use pliers to squeeze the cord restraint where the pedal cord passes through the metal strip on the base and pull it out of the hole.
  1. A basic wire diagram is attached so that you can take note of the color wires and the locations they connect. Remove the white, black, green, and red wires where they connect inside the base.
  1. Pull the original cord out and insert the new through the same hole in the base strip. WARNING! – It is important to also squeeze and return the cord restraint into the hole to prevent damage or injury.
  1. Connect the wires as before.
  1. Test
  1. Relocate & secure the skirt